98
Where:
A = Cross sectional area of the pipe (ft^2)
f = Vortex shedding frequency (pulses / sec)
K
c
= Meter factor corrected for Reynolds Number (pulses / ft)
Q
v
= Volume flow rate (ft^3 / sec)
Q
M
= Mass flow rate (lbm / sec)
V
f
= Flowing velocity (ft / sec)
= Density (lbm / ft^3)
Fluid Calculations
The 240i and 241i density and viscosity calculations and values are determined from Sierra
proprietary fluid properties algorithm based on temperature and pressure of the fluid.