58
Simulation Status:
For the Flow and Frequency menus, use the
key to navigate to the
Simulation Status menu. Push the enter key
to change the status.
Simulate Value
: For the Flow and Frequency menus, use the
key to navigate to the Simulation
Value menu. Push the enter key
and use
to edit the value.
Adjust Value:
For the Minimum Noise Level, Vortex Coefficient Ck, % FS Cutoff, and Output
Averaging menus, the value can be adjusted. Push the enter key
to begin editing. Use
to edit the value. When finished editing, push the enter key
again to save the value.
Autoset Value:
For the Minimum Noise Level and Vortex Coefficient Ck menus, the value can
be Autoset. Use
key to navigate to the Autoset window. Push the enter key
to autoset the
value. The new value will then be displayed.
Default Value:
For the Minimum Noise Level and Vortex Coefficient Ck menus, the value can
be reset to a default value. Use
key to navigate to the Default window. Push the enter key
to reset the value to default. The new value will then be displayed.
K-Factor is a calibration constant and cannot be adjusted.
Simulate Flow
Yes
Sim. Freq.
Yes
Flow ft3/m
100.0
Autoset Ck
21.2
Default Ck
20.0