39
Note that you can also use the function selector of the Smart Interface Portal
(SIP) provided free with the meter.
Table 3: Calibration Sub-Menu
These screens,
Dial-A-Pipe ID, Insertion Depth, Dial-A-Pipe Type
are
available only with the insertion version InnovaMass 241i.
Calibration (Sub-Menu): Gas Type – Dial-A Fluid
Use the “Dial-A-Fluid” sub-menu to change and select the fluid to be measured. Our Dial-A-
Fluid feature is a powerful and unique feature of the 240i/241i. You are able to change your fluid
in the field between air, water, steam, other liquids or other gases .
The display will show fluid type and list the currently selected fluid.
Once on the “Dial-A-Fluid” screen, press enter key
. Press again until the current selected fluid
begins to flash. It may take a few seconds to begin to flash.
Use the up
and down
keys to select between fluid options then hit enter
. The selection
will stop flashing and be permanent.
More fluid choices may be available through Smart Interface Portal (SIP).
You can press the exit
several times to climb back up to the main menu level or press
to
proceed to the next Calibration Sub-Menu called Dial-A-Pipe ID.
Dial-A-Fluid
Carbon Dioxide