53
1.
MCU is malfunctioning.
Solution:
Contact Sierra for remote troubleshooting.
The meter may need to be
2.
MCU has failed.
Solution:
The meter may need to be
Main Board Button Stuck
240i/241i meters have display buttons for navigation and data entry.
Possible Cause of Error Code
1.
Main board button stuck down.
Solution:
Attempt to unstick buttons manually.
Contact Sierra for remote troubleshooting.
Digicomms Fail
240i/241i meters can be ordered with Digital Communications (Hart, Profibus, Foundation
Fieldbus, etc.).
Possible Cause of Error Code
1.
Digital Communications malfunction.
Solution:
Contact Sierra for remote troubleshooting.
The meter may need to be
. If this is not immediately
possible, output via 4-20 mA, RS232, or USB is available.
Signal Noise
240i/241i meters are sensitive to background noise or vibration in the application environment
(60 Hz/50 Hz from power lines).
Possible Cause of Error Code
1.
Meter Settings may need to be adjusted.
Solution: