5. To enable numeric entry <Right Arrow>, then type in the reading of the reference
thermometer (e.g., 72.0).
6. To recalibrate the RTD sensor <ENT>. To verify the calibrated reading, go to the Dual Path
Flow menu [Diagnostic Data/Liquid Data] menu to check the current RTD output. Make
sure that it coincides with the reading of the reference thermometer. Repeat for the other
RTD, if necessary.
Note
Factory Calibration provides an additional prompt after a new temperature is entered: [Are
you Sure? No Yes]. It is recommended that you use [User Cal] to avoid alteration of preset
factory calibration.
Ice Bath RTD Calibration
Use distilled, deionized water and ice mixture at 0°C (32°F) equilibrium for an ice bath. Ensure
temperature with a reference thermometer. Siemens can not assume responsibility for the
incorrect design, construction or operation of an Ice Bath.
CAUTION
Do not allow an RTD sensor to make direct contact with ice during an ice bath calibration
procedure.
To perform a 0°C (32°F) calibration:
1. Immerse RTD sensor in deionized water and ice mixture. Stir the mixture constantly.
2. In the [RTD Calibrate] menu move the highlight by pressing <Up/down Arrow> to the RTD
you want to calibrate (RTD 1 or RTD 2).
3. To access the RTD option list press <Right Arrow>. Move the highlight to [User Cal] then
press <ENT>. This triggers the pop-up window.
4. When you are sure that the RTD Sensor is at 0°C (32°F), press <ENT> to recalibrate the
RTD sensor.
5. To verify the calibrated reading, go to Dual Path Flow [Diagnostic Data/Liquid Data] menu
to check the current RTD output. Make sure that it coincides with the reading of the
reference thermometer. Repeat for the other RTD, if necessary.
7.11
Reflexor
Reflexor is one of the operating modes available on certain SITRANS F 1010 models. The
Reflexor operating mode utilizes Doppler flow detection along with digital signal processing
techniques to successfully measure flow under conditions that may not be suitable for transit-
time flow measurement. The Reflexor samples the flow stream as it detects Doppler shift. It
converts the Doppler shift information by use of Fast Fourier Transform (FFT) and filters the
Functions
7.11 Reflexor
FUP1010 IP67 Portable Flowmeter
96
Operating Instructions, 02/2010, A5E02951522A Revision 01
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......