When you send the command, the flowmeter analyzes the current flow rate for up to sixty
seconds, integrating (averaging) the data for the best zero correlation. During this time, the
menu prompt at the top of the display screen shows a timer that counts from zero to sixty. You
can allow zero averaging for the entire period, or cancel the process at any time by pressing
the <ENT> key. This controls the amount of data the flowmeter averages to obtain a zero level.
Using Reversamatic
If site conditions do not permit stopping the flow rate at the mounting location, and you do not
know the current flow rate, then you can use the ReversaMatic routine to establish the zero
flow level. You should perform the ReversaMatic procedure as quickly as possible to ensure
that the flow rate remains constant throughout the procedure.
To invoke ReversaMatic:
1. To access the [Zero Flow Adjust] option list press <Right Arrow>.
2. Move the cursor to [ReversaMatic]. Press <ENT> to invoke the routine.
3. The flowmeter begins to measure the positive flow rate. "Positive" flow refers to flow moving
from upstream transducer location to the downstream transducer location. Note top prompt
line shows: Reversamatic Action
4. Upon completion, the flowmeter beeps and the display screen shows: Reverse Xdcrs /
Press <ENT>
5. Now remove then remount the upstream and downstream transducers in their reversed
positions. Mount the Up transducer (without removing its cable) in the Down transducer/
cable location. Mount Down transducer with its cable in the Up transducer/cable location.
When remounting the transducers, couple them to the pipe properly. Press <ENT> (after
re-installing the transducers).
6. The flowmeter measures the negative flow rate briefly, then beeps and repeats the prompt:
7. Now remount the transducers for normal operation (in their original orientation). When
remounting transducers, couple them to the pipe properly. Press <ENT> (after re-installing
the transducers).
This completes the ReversaMatic procedure. The system’s zero accuracy will be very close
to that obtainable using the Actual Zero method, providing flow remained constant during this
procedure.
CAUTION
A caution on the use of upper and lower flow limits (used to prevent flow mis-registration)
prior to using the Reversal Zero technique (ReversaMatic): If the negative flow rate that the
flowmeter reads in the step during which the transducers are reversed is more negative than
the lower flow limit, the meter will re-register positive and the Reversal Zero cycle will thus
be corrupted.
Therefore, postpone the installation of upper and lower flow limits until the reversal zero
procedure is executed successfully. For pipes that combine large diameters with very high flow
velocities, it may be necessary to move the upper and lower flow limits out of the way until the
reversal zero is completed. Moreover, pipes of this size frequently have excellent intrinsic zero
performance and may not even need zeroing.
Connecting
5.3 Zero Flow Adjust Menu
FUP1010 IP67 Portable Flowmeter
62
Operating Instructions, 02/2010, A5E02951522A Revision 01
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......