15.Observing the upstream and downstream orientation, attach the UP and DN cables to the
sensors and make snug. Attach the other ends to the UP and DN terminals of the flowmeter
(see figure below).
①
CH 1 Flow Sensor Cable Connectors
③
Channel 1 - Downstream Sensor
②
To Channel 2 Flow Sensor Set
④
Channel 1 - Upstream Sensor
Figure 5-12 Connecting Sensors to Flowmeter
16.Proceed to Commissioning (Page 65).
5.2.5
Direct Mount
Direct Mount - Installation with EZ Clamp and Spacing Guide
The combination of EZ Clamp and spacer guide is the recommended way to mount Direct
Mode sensors. This method ensures that sensors will align exactly 180° from each other and
remain spaced the proper distance apart.
For Direct Mode mounting, a spacer bar is used to establish the distance between sensors
and a spacing guide is used to locate the sensors at the nine o’clock and three o’clock positions.
Should the distance between sensors be beyond the span of a spacer bar, a measuring tape
can be used. The Mylar spacing guide comes in various lengths and widths to accommodate
most pipe sizes.
Spacing Guide P/N
Size
1012-145-1A
2" x 26" (50.8mm x 660.4mm)
1012-145-1
2" x 45" (50.8mm x 1143mm)
1012-145-2
4" x 81" (101.6 mm x 2507.4mm)
1012-145-3
4" x 155 (101.6mm x 3937.0mm)
Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
41
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......