28.Observing the upstream and downstream orientation, attach the UP and DN cables to the
sensors and make snug. Attach the other ends to the UP and DN terminals of the flowmeter
(see figure below).
①
CH 1 Flow Sensor Cable Connectors
③
Channel 1 - Downstream Sensor
②
To Channel 2 Flow Sensor Set
④
Channel 1 - Upstream Sensor
Figure 5-20 Connecting Sensors to Flowmeter
29.Proceed to Commissioning (Page 65).
5.2.6
1012T Mounting Tracks
Using 1012T Sensor Mounting Tracks
The 1012TP and 1012THP Mounting Tracks provide a rigid mounting platform for Series 1011
Universal or high precision size A or B sensors. The mounting tracks service pipe sizes up to
a maximum of 140 mm (5.00") outer diameter. Operating temperatures are supported up to
121°C (250°F). The assembly consists of lightweight aluminium track rails with integral sensor
clamping screws. Attached index pins enable positive locating of the sensors at fixed spacing
locations. Roller-chains and tension screws secure the assembly to the pipe.
The 1012TP mounting tracks support both Direct and Reflect mounting modes. The flowmeter
recommends the appropriate sensors, mounting track and mounting mode, based on the pipe
data entries. Refer to the sensor installation procedures, and if necessary, review how to select
and prepare a pipe mounting location.
Installing a 1012T Mounting Track in Reflect Mode
The Sensor Installation procedures show how the automatic selection of sensors, mounting
mode and spacing method are established. Examine the figure below, which illustrates a typical
Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable Flowmeter
48
Operating Instructions, 02/2010, A5E02951522A Revision 01
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......