16.Wrap the Mylar spacing guide around the pipe so that the left edge is against the sensor
edge mark (see "C" above). Arrange so that one end overlaps the other by at least 8 cm
(3 inches). Trim to fit if necessary, but in order to keep the end square, be sure not to trim
at the overlapping end.
①
Trim material from inner edge if necessary
②
Mylar Spacing Guide
③
8 cm (3-inches) Overlapping Edge
Figure 5-17 Wrapping the Mylar Spacing Guide around the pipe (End View)
17.Realign left edge of the guide with the sensor edge mark. Line up both vertical edges of
the guide and, ensuring that it is snug around the pipe, mark along the overlapping edge.
18.Remove Mylar spacing guide and lay it out on a flat surface. Either measure the exact
distance half-way between the overlap edge and the mark at the overlap, or fold the guide
from the overlap edge to the overlap mark and draw a line at the fold (halfway point).
①
Overlap Edge
③
Mark on Spacing Guide
②
Mark (or fold) exactly at half-way point
④
Circumference
Figure 5-18 Finding the Halfway Distance
19.Reinstall the spacing guide; its left edge abutting the sensor's edge mark on the pipe and
the overlapping edge in line with the dot (now a line) on the pipe (see "C above). Tape it in
this position on the pipe. Take the second frame and place it against the edge of the guide
with its tapered roller centered on the center mark on the guide.
Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable Flowmeter
46
Operating Instructions, 02/2010, A5E02951522A Revision 01
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......