Troubleshooting
10
10.1
Troubleshooting
The following is list of troubleshooting tips and messages that you may encounter. They include
explanations and, in some cases, a recommended action. If a problem seems unsolvable,
contact your local Siemens office or regional Ultrasonic Flow Representative for expert help
at http://www.automation.siemens.com/partner (
http://www.automation.siemens.com/
).
Table 10-1
Troubleshooting Tips
Message
Description
Memory Full!
Response to an attempt to save site data, when data memory is full. Delete an obsolete
site or clear Datalogger memory to make room for the new data.
Memory Corrupted!
Memory read error occurred while accessing the active site data.
Chan Not Setup
Response to an attempt to invoke an operation that requires a channel to be enabled.
Enable the channel [Channel Setup - Channel Enable - Yes]. Note that a channel cannot
be enabled until its transducers are operating.
Clr Active Memory?
Response to pressing the F4 key. Use the F4 function to restore operation if a severe event
(e.g., a violent power surge) disrupts system operation.
Clr Saved Data?
[Clr Saved Data?] only appears after answering No to [Clr Active Memory?]. Answering Yes
to [Clr Saved Data?] will erase ALL saved data.
<EOT>
Response to a request to output Datalogger data to the printer or the Graphics screen when
no Datalogger data exists. Set up the Datalogger.
Empty Pipe - Press <ENT>
Prompt to empty the pipe during the Actual MTY procedure. After emptying pipe, press
<ENT>.
Fill Pipe - Press <ENT>
Prompt to fill the pipe during the Actual MTY procedure. After filling the pipe, press <ENT>.
No Sites - Press <ENT>
Response while trying to recall/delete a site setup when no sites are stored.
Re-space Index
Upon measuring the liquid sonic velocity (Vs), the meter recommends re-spacing the
transducers to improve performance.
Invalid Setup (use Direct
Mode)
During the Initial Makeup the system detects invalid transducer spacing, erroneous liquid
pipe parameters, or some other factor that prevents it from completing the Initial Makeup.
This may be due to one of the following:
● An out-of-range data entry.
● An invalid condition (e.g., overlapping transducers in Reflect Mode). If selecting Direct
Mode does not resolve, review all site setup and transducer installation choices
particularly data entered for pipe and liquid.
● In Reflect Mode the flow meter detects that the pipe wall signal may infringe upon the
liquid signal. Use Direct Mode instead.
● Press <ENT>, <Up Arrow>, <Down Arrow>, or <Left Arrow> to abort install routine.
Continue programming other site data in anticipation of resolving the difficulty later. Call
technical support for help if necessary.
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
117
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......