Table 7-12
Diagnosic Data
Application Information
Function / Explanation
Low Limit
High Pass Filter Setting
High Limit
Low Pass Filter Setting
Noise Level Set
Noise Filter Setting
Doppler Frequency (Hz)
Average FFT Frequency
% Deviation
The width of Spectra is used to determine beam
penetration and calibration. The selection of
"Slurry" in Application Data is recommended for
flow rate indication accuracy when the %
Deviation is less than 25. For % Deviation above
35 the use of "Liquid" is recommended.
For applications that continuously have %
Deviation between 25 and 35 proper selection of
either Liquid or Slurry can be confirmed by
comparison to a known flow reference.
Signal Peak-to-Peak mV
Peak-to-Peak demodulation signal amplitude. The
acceptable range is between 100 and 3200 mV
during typical flow conditions. If less than 100 mV,
check transducer type used and mounting
configuration. If greater than 3200 mV, try in-line
mounting and/or "detune" Carrier Fx.
FFT s/sec
Number of FFTs per second
FFT Peak
Peak FFT Amplitude
Carrier Fx
Transmit Frequency code
Display of "F" at No-Flow conditions
It is normal for the digital display to show an "F" for the no flow condition since the lack of
Spectra information during no flow is the same as a fault condition.
Functions
7.11 Reflexor
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
107
Summary of Contents for SITRANS FUP1010
Page 2: ......
Page 66: ......
Page 136: ......
Page 138: ......
Page 150: ......
Page 153: ...1010WP 7 21614 C ...
Page 154: ...1010WP 7 21614 C ...
Page 155: ...1010WDP 7 21614 C ...
Page 156: ...1010WDP 7 21614 C ...
Page 157: ...1010WP 8 21614 C ...
Page 158: ...1010WDP 8 21614 C ...
Page 159: ......
Page 161: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 162: ...1011PPS 8 TRANSDUCER PART NUMBER L H W NET WT PAIR X Z 21614 C SIZES A B C D SIZE E ...
Page 163: ...21614 C ...
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...21614 C ...
Page 169: ...21614 C ...
Page 170: ...21614 C ...
Page 171: ...21614 C ...
Page 172: ...21614 C ...
Page 173: ...21614 C ...
Page 174: ...21614 C ...
Page 175: ...1015BC 1W 8 21614 C ...
Page 176: ......
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 180: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 181: ...21614 C ...
Page 182: ...1015BC 1 8 21614 C ...
Page 183: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 184: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 190: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 191: ......
Page 192: ......
Page 194: ......
Page 204: ......
Page 206: ......
Page 207: ......