Dosimetry
MOBILETT
SPR8-230.621.01.03.02
97
of 106
Exposure Points
Siemens Exposure Points Table
The table shows how the kV and mAs values translates into exposure points. The
exposure point system is used to easily manipulate film density and contrast.
For example:
Increasing the exposure points one step will yield 25% increased film density.
Increasing the exposure points one step with mAs and decreasing it one step
with kV will yield higher contrast but the same density.
KV
Points
mAs
KV
Points
mAs
KV
Points
mAs
–
-3
0.5
55
8
6,3
96
19
80
–
-2
0.63
57
9
8
102
20
100
–
-1
0.8
60
10
10
109
21
125
40
0
1
63
11
12,5
117
22
160
41
1
1.25
66
12
16
125
23
200
42
2
1.6
70
13
20
133
24
250
44
3
2
73
14
25
–
25
320
46
4
2.5
77
15
32
–
26
400
48
5
3.2
81
16
40
50
6
4
85
17
50
52
7
5
90
18
63
Summary of Contents for MOBILETT XP
Page 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Page 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Page 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Page 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Page 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Page 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Page 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Page 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...