Radiation Calculation
MOBILETT
SPR8-230.621.01.03.02
61
of 106
Calculate the dose area product
The dose area product, DAP, is calculated with
Equation (1)
:
Equation (1)
: DAP = Y x Sx x Sy x Q x k
❏
Y (
µ
Gy x m2/mAs) - conversion factor
❏
Sx - scale index value for collimator adjustment (x-axis)
❏
Sy - scale index value for collimator adjustment (y-axis)
❏
Q (mAs) - quantity of charge during exposure
❏
k (cm2/m2) - constant. k= 6.40 (cm2/m2)
◆
Read the value of Y (
µ
Gy x m2/mAs) from
Diagram 1,
corresponding to the
tube voltage (kV) used during exposure.
◆
Calculate the dose area product, DAP, with
Equation (1)
.