Operation
Operator Manual
SPR8-230.621.01.03.02
34
of 106
Operation
◆
Turn the main switch counterclockwise beyond position (
I
), until the display
switches on. Then let the switch turn back to (
I
).
◆
Pre-position the MOBILETT.
◆
Release the arm system and lift up the tube assembly.
◆
Place the cassette at the required position.
◆
Switch on the collimator light.
◆
Check that the light field is parallel with the sides of the cassette, adjust by
turning the MOBILETT and the collimator.
◆
Set the appropriate film focus distance, using the built-in measuring tape.
◆
Place the detachable protractor on the cassette and read the inclination of the
cassette. Adjust the tube assembly.
◆
Collimate the radiation field for the desired image size and select exposure
parameters.
◆
Release the exposure.
Observe radiation safety regulations,
(→
Page 13)