Technical Data
Operator Manual
SPR8-230.621.01.03.02
82
of 106
DAP-equipment (Option)
Environmental conditions
Technology:
Ionization chamber, Display and control panel
Active area:
146 x 146 mm
Operating temperature:
15 - 40°C
Weight:
0.24 kg
Typical filtration:
< 0.4 mm Al equivalent
Warm-up time:
Ready for use 10 seconds after power on. Specified accuracy achieved 15
minutes after power on.
Chamber light transmis-
sion:
≥
70 %
Energy dependence:
50 - 150 kV
±
8 %, related to 100 kV (IEC 60 580)
40 - 50 kV
±
3 %, related to 50 kV
Reproducibility:
> 99 % under constant pressure and temperature
DAP resolution:
0.1
µ
Gym
2
(0.01
µ
Gym
2
when using a system with high sensitivity)
Maximal measurable DAP: 1 x 10
6
µ
Gym
2
(1 x 10
5
µ
Gym
2
when using a system with high sensitivity)
Dose rate linearity:
< 2 % deviation
Operation
Transport/Storage
Permissible ambient
temperature:
+10°C ... +40°C
-20°C ... +70°C
Permissible relative
humidity:
30 % ... 75 %
10 % ... 100 %
1
1
maximum 15 weeks
Permissible air pressure
area:
700 hPa ... 1060 hPa
Summary of Contents for MOBILETT XP
Page 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Page 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Page 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Page 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Page 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Page 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Page 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Page 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...