Operation
MOBILETT
SPR8-230.621.01.03.02
49
of 106
Exposure with DAP measuring system (Option)
The DAP-equipment is automatically switched on when the main switch of the
MOBILETT is switched on.
◆
Follow normal procedures to release an exposure.
– The measured DAP value will show on the DAP-display in the unit
µ
Gym2
(Micro Gray square meter).
You must reset the DAP value prior to exposure using the Reset button. During
the reset, make sure all segments of the DAP-display light up.
For multiple exposures, the DAP value will accumulate. If DAP values for every
single exposure are wanted, reset the DAP-display by pressing the Reset button
after each exposure.
When the system is switched off, the measured value will be lost.