Dosimetry
MOBILETT
SPR8-230.621.01.03.02
93
of 106
Dosimetry
Exposure Tables
Standard Values
cranial
Chest
Object
Position SID
KV
mAs
Grid
Sensitivity
class
cranial
a.
115
70
8
w.o.
400
cranial
lat.
115
66
6.3
w.o.
400
cranial
axial
115
85
10
w.o.
400
Sinus
a.
115
70
16
w.o.
400
Sinus
lat.
115
70
12.5
w.o.
400
Object
Position SID
KV
mAs
Grid
Sensitivity
class
Lung
a.
130
133
1.25
w.o.
200
Lung/Heart
lat.
130
133
1.6
w.o.
200
Lung, Infant
a.
130
52 - 60
0.63
w.o.
200
Lung/Heart, Infant
lat.
130
60 - 70
0.80
w.o.
200
Ribs 1-7
a.
115
66
4
w.o.
400
Ribs 8-12
a.
115
70
12.5
w.o.
400
Sternum
a.
115
63
10
w.o.
400
Sternum
lat.
115
66
10
w.o.
400
Claviculum
a.
115
60
3.2
w.o.
400
Scapula
a.
115
63
10
w.o.
400
Summary of Contents for MOBILETT XP
Page 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Page 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Page 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Page 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Page 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Page 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Page 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Page 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...