Functional and Safety Checks
MOBILETT
SPR8-230.621.01.03.02
71
of 106
Functional and Safety Checks
To ensure that the system is ready for operation and all functions relevant to
safety are working correctly, you must perform regular function and safety tests.
Daily checks
Please perform the following safety checks daily:
◆
Check the power supply plug. Do not use the MOBILETT if the power supply
plug is damaged.
◆
Check the power cable. Do not use the MOBILETT if the power cable is
damaged.
◆
Visual check of the collimator functions. Switch on the collimator lamp and
check that the light field can be adjusted with the knobs on the collimator.
◆
Check the exposure indicator lamp during exposure.
◆
Check that the audible signal indicates the end of the exposure.
◆
Check the measuring tape regarding function.
◆
Check that the tube assembly and the arm system remain in the desired posi-
tion, but are still easy to move.
◆
Check that the collimator rotates around the beam axis.
◆
For MOBILETT XP Hybrid only: Check charging status of the batteries. If low -
charge to full.
Summary of Contents for MOBILETT XP
Page 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Page 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Page 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Page 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Page 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Page 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Page 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Page 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...