Technical Data
Operator Manual
SPR8-230.621.01.03.02
90
of 106
Guidance and manu-
facturer's declaration -
electromagnetic
immunity
The Mobilett XP is intended for use in the electromagnetic environment specified
below. The customer or the user of the Mobilett XP should assure that it is used
in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications Equip-
ment should be used no closer to any part of the
Mobilett XP, including cables, than the recom-
mended separation distance calculated from the
equation applicable to the frequency of the trans-
mitter.
Recommended separation distance
Conducted RF
IEC 61 000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
d = 1.2 x (
P
)
(1/2)
Radiated RF
IEC 61 000-4-3
3 Vrms
80 MHz to 2.5 GHz
3 V
d = 1.2 x (
P
)
(1/2)
; 80 MHz to 800 MHz
d = 2.3 x (
P
)
(1/2)
; 800 MHz to 2.5 GHz
where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of Equip-
ment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theo-
retically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the location in which the
Mobilett XP is used exceeds the applicable RF compliance level above, the Mobilett XP should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the Mobilett XP.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for MOBILETT XP
Page 34: ...Functional description Operator Manual SPR8 230 621 01 03 02 32 of 106 ...
Page 58: ...Operation Operator Manual SPR8 230 621 01 03 02 56 of 106 ...
Page 60: ...Messages Operator Manual SPR8 230 621 01 03 02 58 of 106 ...
Page 68: ...Radiation Calculation Operator Manual SPR8 230 621 01 03 02 66 of 106 ...
Page 72: ...The Exposure Point System Operator Manual SPR8 230 621 01 03 02 70 of 106 ...
Page 76: ...Functional and Safety Checks Operator Manual SPR8 230 621 01 03 02 74 of 106 ...
Page 100: ...Dosimetry Operator Manual SPR8 230 621 01 03 02 98 of 106 ...
Page 105: ...Index MOBILETT SPR8 230 621 01 03 02 103 of 106 Index ...