Reca
libr
at
ion
1.
Rerun on-line calibration to verify the deviation: press
to return to P358.
2.
Verify the mechanics of the flowmeter: carry out material tests to ensure the
readings are correct. (See page 30,)
3.
If the mechanics are functioning correctly, perform an initial span using P388. (See
Factoring
To calculate the value of a new or unknown test weight to the current span, use the
factoring procedure.
With the material flow turned off:
Access P359 in VIEW mode
Press
Press
Press
Note:
For optimum accuracy in the factoring results, a routine zero calibration is
recommended prior to performing the factoring routine.
P359 Factoring
V
Enter 1 to start factoring
1
Factoring Weight
Place weight and press ENTER.
Factoring Weight
Factoring in progress
##.##
Factoring Weight
Press ENTER to accept value
45.25
P359 Factoring:
V
Enter 1 to start factoring
1
Note:
If multispan function is used, the test rate value is stored for the current
multispan only.
e.g. factor the test weight
the rate reported while factoring is in
progress.
e.g. the new factor, if accepted