7ML19985CN01
Milltronics SF 500 - INSTRUCTION MANUAL
Page 25
m
mmmm
St
art
Up
The test rate value should be less than the design rate. If not, contact Siemens
Milltronics.
The initial programming requirements are now satisfied. To ensure proper entry of all
critical parameter values, return to P002 and review parameters through to P017.
Load Cell Balancing
If you are operating a two-load cell solid flowmeter, balance the load cells electronically.
Do this prior to initial programming and calibration, or after either or both load cells have
been reinstalled or replaced.
Unbalanced load cells adversely affect the performance of your solids flowmeter
weighing system.
Typical two load cell flowmeter
P017 Test rate: Weight MS 1
V
Enter test rate
75 t/h
C
load cell
’B’
test weight connection
points
load cell
’A’