Milltronics SF 500 – INSTRUCTION MANUAL
7ML19985CN01
mmmm
m
Recalibr
at
ion
The SF 500 On-line material total, the amount of material totalized between the High and
Low limits, is compared to the value entered in P356. The deviation percentage between
these values and the new Span count value is displayed.
Press
to accept the results.
If you want to reject the results and return to RUN mode, press
.
If you want to reject the results and perform another on-line calibration, press
to
return to P358.
Access
Press
If the deviation is greater than
±
12%:
On-line Calibration
Deviation
2.51%
Press ENTER to accept
New span
22280
On-line Calibration Complete
Press ENTER to accept
New span 22280
Note:
•
Deviation must be no greater than
±
12% of the initial span or it will not be
accepted.
•
For remote access, On-line Calibration can be accepted using one of the auxiliary
inputs (refer to
P270 Auxiliary Input Function
on page 100).
Rate
0.00 t/h
Total 1:
10.15 t
Note:
For remote access, to return to RUN mode, program one of the auxiliary inputs
(refer to
P270 Auxiliary Input Function
P358 On-line Calibration Features
V
0-OFF,
1-ACTIVE
1
Calibration is out of range
Deviation Error:
deviation percent
new Span count value