Page 28
Milltronics SF 500 - INSTRUCTION MANUAL
7ML19985CN01
mmmmm
St
art U
p
Zero Calibration
Press
Press
The duration of the Zero calibration is dependent upon the present time duration and the
(P360) calibration duration.
Press
Press
Accepting the Zero returns to start of Zero. Perform a new Zero, or continue to Span.
Span Calibration
When performing a Span Calibration where the test reference is ECal (P002 = 2), do not
apply the supplied test weight and run the flowmeter empty
.
Stop the material flow and apply the test weight to the flowmeter as instructed in the
flowmeter manuals.
Press
Note:
To obtain an accurate and successful calibration, ensure that no material is
flowing through the flowmeter and that the test weights are not applied.
Zero Calibration: Current Zero
0
Clear flowmeter. Press ENTER to Start
Initial Zero Calibration. In progress
Current Reading:
#####
Calibration Complete. Deviation
0.00
Press ENTER to accept value:
551205
Zero Calibration. Current Zero
551205
Clear flowmeter. Press ENTER to Start
Note:
To obtain an accurate and successful calibration, ensure that there is no
material flowing through the flowmeter and that the test weight is applied.
Span Calibration. Current Span
0
Setup test. Press ENTER to Start
the current zero count
the zero count being calculated
while calibration is in progress
for example, the new zero count, if
accepted
the deviation from previous zero. For
an initial zero there is no previous
zero; hence the deviation is 0.
the current zero count of 551205
the current span count