Operation
5.8 Calibration
FIDAMAT 6-G
5-22
Operating instructions, 07/2006, A5E00478463
Settings for AUTOCAL Check
Figure 5-20 Autocal check
In the subfunction
"Autocal Check", enter the desired calibration tolerances which should be
checked by Autocal Check. With functions 71 and 72, you can also activate the relay output
or binary input for "Autocal Check".
When a calibration limit is exceeded, the maintenance request W10 and, if parameterized,
the relay "Acal Chk Dif" are set.
Both will be reset again after an error-free Autocal Check. In the logbook, W10 is retained.
"For starting via cycle, initiate Autocal Check":
If this function is activated, the device executes an Autocal Check when Autocal is started
via a cycle. In other words, it only checks that the calibration tolerances are kept, but doesn't
carry out calibration.
Sequence:
1.
Start the Autocal Check:
–
by means of the softkey
"Initiate Autocal Check once" in the subfunction "Acal/-Check
sequence"
–
via binary input
–
via cycle
2.
The device executes the sequence as parameterized in the subfunction
"Autocal/Check
sequence".
See also
Warm-up phase and operating states (Page 4-5)
Relay allocation (function 71) (Page 5-40)
Binary inputs (function 72) (Page 5-42)