Operation
5.11 Configuration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-53
5.11.10
Analog output memory (function 77)
77 Store THC
analog out.to meas.value
analog out.to O/2/4 mA
analog out.to 21 mA
Store on/off
Figure 5-56 Setting the analog output
With this function, you set which current is output at the analog output when a fault occurs or
when the functional check "CTRL" is activated (e.g. during calibration).
With softkey 5, you can activate (■) or deactivate (□) the entire function.
Switch softkey 1 to active (■) in order to output the last measured value at the analog output.
Switch softkey 2 to active (■) in order to output 0/2 or 4 mA.
Switch softkey 3 to active (■) in order to output 21 mA.
5.11.11
Calibration tolerances (function 78)
Setting the calibration tolerances
78 Calib. tolerance THC
Calib. tolerance at zero
in % of smallest MR: :1O:
Calib. tolerance of sens.
in % of current MR: 1O
Signal tolerance
voilation.
Figure 5-57 Setting the calibration tolerances
With this function, you define the calibration tolerances.
With softkey 3
"Flag if tolerance is exceeded", you activate or deactivate the tolerance
monitoring.