Commissioning
4.2 Commissioning preparations
FIDAMAT 6-G
4-2
Operating instructions, 07/2006, A5E00478463
Warning
Toxic and/or corrosive gases
However, this does not apply to toxic or corrosive gases! In this regard the maximum
allowable workplace concentration figures must be observed. With such gases, flushing must
always be carried out after work is completed.
4.2
Commissioning preparations
4.2.1
General information for commissioning
Device position
The FIDAMAT 6 may only be operated in its horizontal position.
Operation
Before connecting and switching on the device, the operator must have made himself
familiar with the device operation.
Interfaces
Before use, the interfaces must be correctly allocated and parameterized.
Noise attenuation
The output signal fluctuations caused by noise in the measuring signal can be reduced with
function 50. This function offers you the option of parameterizing a low-pass filter which can
be assigned a time constant up to 300 s.
Temperature influence
During operation, make sure that the permissible ambient temperature of 5°C to 45°C is
kept.
See also
Connection of the signal lines (Page 3-8)
Technical data of the FIDAMAT 6 series (Page 7-1)
Technical data of the FIDAMAT 6-G (Page 7-3)