Commissioning
4.3 Commissioning and operation
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
4-7
Note
State changes are caused either manually using function 91, or automatically by occurring
faults / maintenance requests.
See also
State changes due to fault / maintenance request (Page 4-7)
Set external pressure (function 62) (Page 5-37)
Start-up state (function 91) (Page 5-62)
Pressure values (function 92) (Page 5-63)
4.3.3
State changes due to fault / maintenance request
In the following tables, you will find potential automatic state changes of the device:
Table 4-1
Change in state due to fault
Fault
Last state
New state
S1: Parameter memory
- - -
S2: Pump doesn't start up
(only with FIDAMAT 6-E)
Warm-up | Standby
Standby
S3: Flame doesn't ignite
Warm up
Pause
S4: External fault
- - -
S5: Oven temperature
Warm-up | Standby | Measure |
Pause
Pause without heating
S7: Flame temperature
Warm-up | Standby | Measure |
Pause
Pause
S8: Combustion air/gas pressure
Warm-up | Standby | Measure |
Pause
Pause
S9: Heater is off
Warm up
Pause without heating
S10: 24 h RAM/flash check
- - -
S12: Mains voltage supply
Warm-up | Standby | Measure |
Pause
Pause
S13: Hardware / mains frequency
Warm-up | Standby | Measure |
Pause
Pause
S14: Measured value >
Characteristic full-scale value
- - -
S15: Calibration terminated
- - -