Operation
5.11 Configuration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-63
5.11.21
Pressure values (function 92)
Figure 5-70 Changing the pressure values
In this function, various pressure values are displayed. You can change some of these:
Combustion gas pressure
Parameters
Explanation
Nominal pressure
is automatically displayed
max. deviation
If the combustion gas pressure deviates from this
nominal pressure value, the device exits the
"measure" operating state. This value can be
changed.
Actual pressure
is automatically displayed
Combustion air pressure
Parameters
Explanation
Nominal pressure
is automatically displayed
max. deviation
If the combustion air pressure deviates from this
nominal pressure value, the device exits the
"measure" operating state. This value can be
changed.
falling short to standby (min:):
If the combustion air pressure falls below the set
pressure level by the set value, the unit switches
over to the "standby" operating state.
Actual pressure
is automatically displayed