132
DCA Vantage Operator’s Guide
Tro
ubles
hoot
in
g
E26
Excessive
noise in sample
reading
The standard deviation
of the 16 readings at the
Sample channel during a
reading taken is too
large in Sample read
position.
Possible causes:
•
low lamp intensity
•
lamp alignment/
rotation
•
lamp intensity
variation
1. Discard the sample.
2. Power off the system.
3. Clean the cartridge area. Refer
Cartridge Spring and Cartridge
Area
4. Restart the system.
E27
Excessive
Lamp Drift
The change in mean
signal between
successive Air readings
is too large at either the
Sample channel or the
Reference channel.
Possible causes:
•
dust in the optical
path
•
lamp filament
orientation
1. Discard the sample.
2. Power off the system.
3. Clean the cartridge area. Refer
Cartridge Spring and Cartridge
Area
4. Restart the system.
E30
Thermal control
system error -
low
The temperature
measured by one of the
cartridge holder
thermistors is
2
º
C
.
Possible causes:
•
defective heater
plate
•
defective thermistor
assembly
•
defective connector
•
defective electrical
circuit
1. Discard the sample.
2. Restart the system.
Error Code
Description
Corrective Action
Summary of Contents for DCA Vantage
Page 10: ...10 DCA Vantage Operator s Guide ...
Page 24: ...24 DCA Vantage Operator s Guide ...
Page 40: ...40 DCA Vantage Operator s Guide ...
Page 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Page 120: ...120 DCA Vantage Operator s Guide ...
Page 150: ...150 DCA Vantage Operator s Guide File Management ...
Page 166: ...166 DCA Vantage Operator s Guide ...
Page 192: ...192 DCA Vantage Operator s Guide ...