50
DCA Vantage Operator’s Guide
Ope
rat
ing the
Sy
st
em
Filling the Glass Capillary with Blood from Finger Stick
When the capillary is filled with the sample, analysis must begin within 5
minutes.
1. Hold the capillary holder at an angle.
2. Touch only the tip of the capillary to a small drop of blood on the finger until
the capillary fills.
NOTE:
1
µ
L of blood is required to fill the capillary.
Figure 10 Filling Capillary Holder with Blood from Finger Stick
Filling Glass Capillary with Blood Obtained by Venipuncture
When the capillary is filled with the sample, analysis must begin within 5
minutes.
1. Mix the sample well (by inversion or use of a tube mixer) to prevent
separation of red blood cells and plasma.
2. Remove stopper from blood collection tube in such a way that a small
sample of blood remains on the stopper.
3. Hold the capillary holder at an angle.
4. Touch only the tip of the capillary to the blood sample on the stopper.
NOTE:
Do not attempt to fill the capillary by touching the glass capillary to
blood in a blood collection tube. Attempting to fill the capillary in this manner
most often results in blood touching the capillary holder. If blood touches the
capillary holder, discard the capillary holder.
1
Glass Capillary Tip
Summary of Contents for DCA Vantage
Page 10: ...10 DCA Vantage Operator s Guide ...
Page 24: ...24 DCA Vantage Operator s Guide ...
Page 40: ...40 DCA Vantage Operator s Guide ...
Page 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Page 120: ...120 DCA Vantage Operator s Guide ...
Page 150: ...150 DCA Vantage Operator s Guide File Management ...
Page 166: ...166 DCA Vantage Operator s Guide ...
Page 192: ...192 DCA Vantage Operator s Guide ...