DCA Vantage Operator’s Guide
133
Tr
oubles
hoot
ing
E31
Cartridge
temperature is
very low
The temperature
measured by one of the
cartridge holder
thermistors is
< (T
setpoint
- 12.9)
º
C
while the system is idle,
during an Optical Test,
during the first 5 minutes
of an HbA
1c
test, or the
first 5 minutes of a
Microalbumin/Creatinine
test.
Possible causes:
•
cold cartridge
•
defective heater
plate
•
defective thermistor
assembly
•
defective connector
•
defective electrical
circuit
1. Ensure the reagent cartridge
was properly prepared for
testing. Reagents should be at
room temperature for testing.
2. Power off the system.
3. Rerun the test.
E32
Cartridge
temperature
too low during
measurement
The temperature
measured by one of the
cartridge holder
thermistors.
Possible causes:
•
cold cartridge
•
defective heater
plate
•
defective thermistor
assembly
•
defective connector
•
defective electrical
circuit
1. Ensure the reagent cartridge
was properly prepared for
testing. Reagents should be at
room temperature for testing.
2. Power off the system.
3. Rerun the test.
Error Code
Description
Corrective Action
Summary of Contents for DCA Vantage
Page 10: ...10 DCA Vantage Operator s Guide ...
Page 24: ...24 DCA Vantage Operator s Guide ...
Page 40: ...40 DCA Vantage Operator s Guide ...
Page 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Page 120: ...120 DCA Vantage Operator s Guide ...
Page 150: ...150 DCA Vantage Operator s Guide File Management ...
Page 166: ...166 DCA Vantage Operator s Guide ...
Page 192: ...192 DCA Vantage Operator s Guide ...