DCA Vantage Operator’s Guide
203
2. Select the type of connection,
Ethernet
or
Serial
, and the protocol,
ASTM
or
POCT1A
.
3. To enable Auto Send, select the
Automatically Send
check box.
This selection is disabled by default.
To enable manually sending selected results, clear this check box.
4. Select
Configure
to set up the Ethernet or Serial port connection.
NOTE:
At this stage, the system does not verify whether a connection
exists.
5. Continue configuring the Ethernet or Serial connection parameters.
6. When finished, select
Save
.
The Auto Send setting is included in the backup and restore data so an
administrator can copy the settings from one DCA system to another.
Transmission Status
The system displays the status of the transmission on the test result screen
when communications are enabled, so that you know whether the LIS has
received the transmission.
The transmission status can be one of the following:
•
Not Sent: The message has not been transmitted to the LIS.
•
Sending: Transmission is in process, but the system has not received
acknowledgement from the LIS. The connection might have gone down
either during or after the attempted transmission, and the connection has
not been restored.
For the Ethernet ASTM protocol, the system tries again to transmit the
results if the connection is restored within 72 hours of the most recent
successful transmission. For the POCT1-A protocol, no time limit exists for
retries.
•
Received: The message was sent to the LIS, and the LIS has
acknowledged receipt of the message.
•
Failed: The LIS did not receive the message, or the LIS did not
acknowledge the receipt of the message.
•
For a serial connection using the ASTM protocol, the test result has
failed to be received by the LIS on the first attempt to transmit.
•
For an Ethernet connection using the ASTM protocol, 72 hours have
passed since the most recent manual-send or auto-send after a good
connection went down, and the test result has failed to be received by
the LIS on subsequent attempts to transmit.
•
Unknown: Only for test results created on Version 2.0 and upgraded
directly to version 4.0 or greater. Not for live tests.
Summary of Contents for DCA Vantage
Page 10: ...10 DCA Vantage Operator s Guide ...
Page 24: ...24 DCA Vantage Operator s Guide ...
Page 40: ...40 DCA Vantage Operator s Guide ...
Page 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Page 120: ...120 DCA Vantage Operator s Guide ...
Page 150: ...150 DCA Vantage Operator s Guide File Management ...
Page 166: ...166 DCA Vantage Operator s Guide ...
Page 192: ...192 DCA Vantage Operator s Guide ...