146
DCA Vantage Operator’s Guide
Tro
ubles
hoot
in
g
E315
Microalbumin
normalization
required
The Microalbumin
Normalization was not
performed or the
Microalbumin
Normalization Factor
(MNF) was corrupted.
Possible cause:
•
system not adjusted
to run Microalbumin/
Creatinine test
1. Acknowledge the error.
2. If the error still occurs, contact
your local technical support.
E316
Microalbumin
normalization
out of range
The Microalbumin
Normalization Factor
entered is outside of
expected range.
Possible causes:
•
normalization
cartridge is not in
proper position
•
normalization
cartridge is dirty or
defective
1. Acknowledge the error.
2. Verify that the cartridge was
inserted completely into the
system.
3. Repeat the test.
4. If the error still occurs, contact
your local technical support.
Error Code
Description
Corrective Action