126
DCA Vantage Operator’s Guide
Tro
ubles
hoot
in
g
Barcode cannot
be read using the
Onboard barcode
reader
(continued)
Ensure the
barcode is facing
the correct
direction.
Refer to
Is the label
smudged or
contains extra
marks?
Verify label is clean and clear of
extra markings.
Does the problem
only occur when
scanning the
calibration card?
The Calibration card must be held
flat against the scanner slot or it will
not read the barcode.
Does the problem
only occur when
scanning the
reagent barcode?
Document the issue with the
reagent lot and replace the reagent.
Contact your local support provider.
Did you scan the
same side of the
Microalbumin or
Creatinine card
twice?
When calibrating Microalbumin or
Creatinine, after scanning side 1,
the system prompts you to scan
side 2. If you scan side 1 again, the
system ignores the scan.
Buffer Tab Hard
to Pull
If you pull the foil
tab too fast, it can
tear the foil and
impede the full
buffer release
Slowly pull the buffer tab while
anchoring the cartridge with 2
fingers.
Calibration Lot
Expired
Check the date
on the control
card. If the date is
valid, the
DCA Vantage
system may be
set with an
incorrect date.
Refer to
Cartridge is stuck
in the instrument
If the cartridge tab has broken off,
do not try to remove the cartridge.
Contact your local technical support
provider.
Symptom
Possible Cause
Action
Summary of Contents for DCA Vantage
Page 10: ...10 DCA Vantage Operator s Guide ...
Page 24: ...24 DCA Vantage Operator s Guide ...
Page 40: ...40 DCA Vantage Operator s Guide ...
Page 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Page 120: ...120 DCA Vantage Operator s Guide ...
Page 150: ...150 DCA Vantage Operator s Guide File Management ...
Page 166: ...166 DCA Vantage Operator s Guide ...
Page 192: ...192 DCA Vantage Operator s Guide ...