62
DCA Vantage Operator’s Guide
Ope
rat
ing the
Sy
st
em
A supervisor (that is, a Level 1 operator) can designate the required primary
and optional secondary HbA
1c
reporting units for printing data. Based on this
setting, the system displays the appropriate primary, and optional secondary
HbA
1c
reporting units. See
Designating the Primary and Secondary HbA1c
NOTE:
You can print, enter, or edit the sample data while the results are
displayed.
Removing the Reagent Cartridge
When the tests are complete, remove the reagent cartridge from the system.
1. Open the cartridge compartment door.
2. Locate the button on the right side of the cartridge compartment.
3. Push and hold it down with your right hand.
4. With your left hand, gently push the tab on the cartridge to the right.
This action releases (unlocks) the cartridge.
Figure 16 Removing the Reagent Cartridge
1
Reagent Cartridge
2
Release Button
Summary of Contents for DCA Vantage