98
DCA Vantage Operator’s Guide
Ma
in
te
n
a
nc
e
Removing and Cleaning the Cartridge Spring and Cartridge Area
Materials Required:
•
lint-free cloth
•
water or ethanol
•
paper clip or similar device
•
mild detergent
•
sponge swab
1. Ensure the power is off.
2. Open the cartridge compartment door as far as possible.
3. Wipe the inside surface of the compartment door and surfaces on both
sides of the compartment using a lint-free cloth dampened with water or
ethanol.
4. Dry the surface using a clean, dry, lint-free cloth.
5. Locate the cartridge return spring inside the cartridge holder.
WARNING
Turn the power off and disconnect the power cord before cleaning
the cartridge compartment.
CAUTION
Do not allow liquid to drip into system. If liquid drips into the system,
you can damage the optics.
BIOHAZARD
Wear personal protective equipment. Use universal precautions.
Refer to
‚ page 167, for recommended
precautions when working with biohazardous materials.