
4. If necessary, the transducer cables can be shortened at the respective transducer / sensor end
(the high temperature versions).
Make sure the cables are equal in length in order to avoid signal delays in the signal
processing.
Note
The SITRANS F US sensor is shipped with 2 or 4 transducer cables of fixed lengths of 3, 15, 30,
60, 90 or 120 m (9.84, 49.21, 98.43, 196.85, 295.28, or 393.70 ft).
Note
The allowed minimum transducer cable bending radius is 50 mm.
5. Tighten the cable glands on the threaded bush until the cable is sealed tightly (IP65).
6. Screw the lid back on.
7. After the installation, check and, if necessary, correct the cable length setting of the
transmitter (see menu 7).
Wiring 1-path systems
1
)
①
For 1-path sensors exchange the two unused cable glands with the blind plugs.
Figure 5-2
Wiring transducer cables, 1-path system.
Connecting
5.1 Wiring the transducer cables
SITRANS FUS060 with PROFIBUS PA
34
Operating Instructions, 09/2021, A5E02225054-AD