Menu
code
Device function,
Parameters
Description
Factory setting
Setting options
Read
or write
2.2
Suppress error
Switch off error message
None
None
Unreliable flow
Measuring path dis‐
turbed
Both
2.2
Suppress error
(not stored)
Write
2.3
Device test
Test device status
"OK"
2.3.1
Self test
Visual LCD test
Read
2.3.2
Display test
Read
2.4
Simulation
2.4.1
Flow Value Time
Simulation of flow.
Simulation value. Affects all outputs
and totalizers.
Duration of simulation. After expira‐
tion of this time the normal measur‐
ing mode is resumed.
0%
End (no action)
-110 to +110%
End (no action)
0 min
30 min
60 min
Write
2.4.2
Digital output 1
Simulation of output signal
End (no simulation) End (no simulation)
0.1 Hz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
Alarm on
Alarm off
Write
D.1.3
Menu 3 - Measuring functions
Menu
code
Device function,
Parameters
Description
Factory setting
Setting options
Read
or write
3.1
Flow
3.1.1
Unit
Physical units for volume flow, mass
flow
Note: Whenever using mass flow -ac‐
tual density have to be selected in
menu 3.1.4
m
3
/h
l/s, l/min, l/h, m3/s,
m3/h, m3/d, Ml/d,
ft3/s, ft3/min, ft3/d,
gal/s,
gal/min, gal/d,
Mgal/d, ImpGal/s, Im‐
pGal/min, ImpGal/d,
g/s, g/min, g/h, kg/s,
kg/min, kg/d, Ton/
min,
Ton/h, Ton/d, lb/s, lb/
min, lb/h, lb/d, STon/
min, STon/h, STon/d
HMI menu structure
D.1 Menu items (FW Rev. 4.00.00)
SITRANS FUS060 with PROFIBUS PA
108
Operating Instructions, 09/2021, A5E02225054-AD