O P E R A T I N G I N S T R U C T I O N S | VICOTEC320
Subject to change without notice
MAINTENANCE
9
Fig. 37: Sender lamp with LED unit
Exchanging the sender lamp with LED unit
1
Loosen 6 screws on the rear side of the SR-unit and swivel the rear side out.
2
Pull off the voltage supply line of the LED.
3
Loosen screw (crosshead) of the plug of the sender lamp voltage supply and disconnect
the plug.
Fig. 38: Sender lamp
1 Sender lamp
2 LED unit
3 Connection of LED unit power supply
4 Connection of sender lamp power supply
WARNING: Hazard by voltage.
Live parts are accessible when the device is open!
▸
Switch the supply voltage off before opening the device.
▸
Only use suitable, insulated tools.
1 Connection of voltage supply line
2 Screws of sender lamp
3 Connection of LED unit power supply