USING THE VICOTEC320
Operating mode switch
Fig. 32: Menu Maintenance/Operating mode switch
This screen serves to switch between Measuring mode and Maintenance mode. Apart from
that, a reference cycle and a system restart can also be initiated. The parameters are not
deleted. The connection between SOPAS and VICOTEC320 must be established again after
a system restart (
see “Creating a connection”, page 38
8.2.3
Saving the current parameter set
1
Save the parameter set. The saved file can then be restored, for example on new hard-
ware.
Select:
Device/Export SDV file
2
Storing the project (a “project” can be several devices). This file can then, for example,
be printed but can however
not
be restored in the device.
a) To store the current parameter set, select the
Save as
command in the
Project
menu.
b) Enter a file name in the dialog window and confirm with [Save].