44
8011703/XF49/2019-09| SICK
O P E R A T I N G I N S T R U C T I O N S | VICOTEC320
Subject to change without notice
8
USING THE VICOTEC320
8.2.2
Reading out the VICOTEC320 and operating manually
The corresponding operator level must first be selected to configure a device with the
SOPAS ET software. The SOPAS ET software runs in the operator level “Operator” after com-
missioning and parameters can only be read.
– To change the operating level to “Authorized operator”: Login via the [Login] button >
enter password (
– Double-click one of the functions in the project tree to start this function.
– To save all the data, select the “Export SDV file” command in the
Device
menu .
The following tabs are important for you; the other tabs are shown colored gray and are only
relevant for Service technicians.
Bargraph measured values
Fig. 29: Bargraph measured values
This screen shows whether the sensors are in measuring operation or whether a fault or
maintenance requirement exist.
Apart from that, the current measured values for visibility, temperature, NO, NO
2
and NO
x
are displayed (depending on the device variant).
When fault or maintenance request is shown, the measurement triggering the fault or
maintenance request is shown next to the measured values.
The “Uncertain” LED next to the measured values signals that the measured value is
“uncertain” (e.g.: Calibration range exceeded.
→
Logbook).