35
8011703/XF49/V3-1/2019-09| SICK
O P E R A T I N G I N S T R U C T I O N S | VICOTEC320
Subject to change without notice
COMMISSIONING
6
6.4
Spectral analysis
Fig. 24: Menu Parameter/Spectral analysis
1
Open “Parameter/Spectral analysis” in the SOPAS menu tree.
2
Enter the installation altitude (above sea level).
6.5
Visibility and smoke detection
Fig. 25: Menu Parameter/Visibility
1
Open “Parameter/Visibility” in the SOPAS menu tree.
>>> Default setting: Smoke detection not activated.
2
To activate smoke detection: Set checkmark.
Measured values of gases are kept constant at first when the light beams are
interrupted. A malfunction message is sent to the evaluation unit when interruptions
last longer than two minutes.
A value above 22 mA is transmitted at the analog output when the light beams are
interrupted (quick drop of visibility). No malfunction message is output.