MAINTENANCE
9
9
Maintenance
9.1
Cleaning
Set device to Maintenance mode
9.1.1
Cleaning sensors
The tube can be removed in order to clean the protective screen of the sender/receiver unit
or reflector.
1
Loosen both nuts at the end of the tube and pull the tube off.
2
Clean the protective screens with clean optical tissues.
3
Check the tube and the optical beam path for soiling though deposits or animals and
clean when necessary.
4
Position the tube and tighten both nuts.
9.1.2
Tunnel cleaning
▸
Set device to Maintenance mode
Fig. 33: Menu Maintenance/Operating mode switch
1
Open
“Maintenance/Operating mode switch”
in the SOPAS menu.
2
Click [Maintenance] below Operation.
CAUTION: Eye injuries caused by improper handling of UV or blue light beams
The UV beam of the deuterium lamp or the blue light beam of the LED can cause serious
injuries through direct eye and skin contact. This results in the following safety mea-
sures when working on the switched on device with access to the light beam exit:
▸
Always wear UV protective goggles (in accordance with standard EN 170)
▸
The UV goggles do not provide protection against injuries caused by blue light beams,
therefore switch the LED off when working.
▸
Use the lamps only when they are perfectly safe. Operation is not allowed when the
lamp, supply lines or operating parts are visibly damaged.
Cover every sensor tube with a protective cap during tunnel cleaning.