INSTALLATION
5
4
Screw the wall holder on and tighten the screws with 70 Nm according to the
manufacturer's assembly specifications, use a torque wrench as necessary.
5
Screw the consoles on provisionally at first.
Fig. 10: Wall holder assembly drilling plan
5.6.2
Fitting the VICOTEC320 sensors
1
Position the sender/receiver unit on the
left
assembly console and screw it on lightly
with both the retaining screws included in the delivery.
2
Position the reflector on the
right
assembly console and screw it on lightly with both the
retaining screws included in the delivery.
3
Screw a laser adjustment unit on each sensor above the tube using both knurled-head
screws.
Fig. 11: Laser adjustment unit
4
Switch on the laser adjustment unit on the reflector.
Tolerances for measuring path A
10
±0.1
m
105
105
210
210
A
182
182
15
0
15
0
15
0
15
0
15
0
15 x 50
+2
0
15
0
1
Crosshair
2
Knurled-head screws
3
Adjustment laser
WARNING: Laser class 2
▸
Laser radiation – do not look in the beam.