INSTALLATION
Not included in scope of delivery:
●
Stainless steel assembly consoles for the sensors
5.6
Assembly
Assembly work must only be carried out by skilled persons familiar with the assembly work.
5.6.1
Fitting the assembly consoles
The assembly consoles comprise 3 parts:
Fig. 9: Assembly console
Two angle bracket screws fasten the wall holder and the assembly console together. The
angle bracket can be swiveled up to ±7.5° to compensate any assembly unevenness.
Procedure
1
Determine the installation location for the assembly consoles according to the project
planning.
2
Drill the wall holder openings according to the Drilling plan, see
assembly drilling plan”, page 21
3
Insert dowels or wall ties according to the manufacturer's assembly specifications (walls
must be made of at least C20/25 to C25/30 concrete).
1
Wall holder for wall fitting
2
Angle bracket to fasten the sensor
3
Angle bracket screws
Observe the following points during assembly:
▸
Observe local valid safety measures.
▸
Only use high-strength and absolutely non-corrosive fastening material made of
stainless steel because the tunnel atmosphere is highly corrosive.
▸
Ensure there is enough space to remove the tube and device cover.
▸
Keep the length of the measuring section as exact as possible. Record small
deviations in the Assembly protocol.
▸
Mount both assembly consoles at the same height. Height differences in the optical
axis can be compensated later by swiveling the angle bracket.
▸
Align both assembly consoles at the same tilt angle to the tunnel wall. Different tilt
positions to the tunnel profile make the following sensor alignment difficult. Insert
washers under the wall holder when necessary.