O P E R A T I N G I N S T R U C T I O N S | T - E A S I C
®
8023072/1E22/2022-02-23 | SICK AG
Subject to change without notice
1 6
5
MOUNTING
At least 5-10DN
At least 5-10DN
At least 3-5DN
At least 3-5DN
Fig. 4: Recommended inlet and outlet zones
For vertically installed pipes, installation in the riser is recommended. The sensor must
not be installed in front of or in a downpipe, or in an open downpipe.
Fig. 5: Vertical installation position
For horizontal pipes, side mounting or mounting from below is possible (only applies if
the pipe is free of deposits).
Fig. 6: Horizontal installation position
Mounting from above is only possible if the pipeline is completely filled with medium.