O P E R A T I N G I N S T R U C T I O N S | T - E A S I C
®
8023072/1E22/2022-02-23 | SICK AG
Subject to change without notice
4 4
7
COMMISSIONING
NOTE
The analog output outputs the current that was parameterized for Q=0L/min.
The pulse output does not output any pulses at Q=0L/min.
Cut-off value + hysteresis
Cut-off value
Flow
Zeit
Low flow suppression active
Switch-on point for low flow suppression
Hysteresis in percent
(5 ... 300% of the cut-off value)
Switch-off point for low flow suppression
Measured flow
Displayed flow
Fig. 33:
Low flow cutoff function
7.4.10
Delay Mode Q1
This parameter is used to define the delay mode for Q1. The following parameters are
available for selection:
Parameter
Function
Off
No delay on Q1 status change.
Ton Delay
When
Q1 Switchpoint 1
is reached, the status of Q1
does not change until
Delay time Q1
Toff Delay
When
Q1 Switchpoint 2
is reached, the status of Q1
does not change until
Delay time Q1
Ton&Toff Delay
When
Q1 Switchpoint 1
is reached, the status of Q1
does not change until
Delay time Q1
) has elapsed. When
Q1 Switchpoint 2
is
reached, the status of Q1 does not change until
Delay
time Q1
(see