Access
Command
Name
Parameter
Com‐
ments
Channel A Channel B Channel C Channel D
R/W
0x0208
0x0288
0x0308
0x0388
Peak num‐
ber
0x00: No.
1 peak to
0x07: No.
8 peak
R/W
0x020C
0x028C
0x030C
0x038C
Peak num‐
ber 2 (for
thickness
measure‐
ment)
0x00: No.
1 peak to
0x07: No.
8 peak
Can only
be used
when
thickness
measure‐
ment is
selected
R/W
0x0220
0x02A0
0x0320
0x03A0
Measure‐
ment off‐
set
*5
R/W
0x0234
0x02B4
0x0334
0x03B4
Execute
teach
0x00:
None
0x01: Exe‐
cute offset
0x02:
Clear off‐
set
Also used
to exe‐
cute span
teach
Is
returned
in
response
to read
requests
when
teach is
being exe‐
cuted
*5: Explanation of the parameters (measurement offset)
The applied parameter (in nm) is added to the raw displacement measurement. Alter‐
natively, after a zero set operation has been executed, this parameter contains the cal‐
culated offset.
7.4.3.2
Making the settings for channels A–D
Table 37: Command list for the settings for channels A–D
Access
Command
Name
Parameter
Com‐
ments
Channel A Channel B Channel C Channel D
R/W
0x0210
0x0290
0x0310
0x0390
Median fil‐
ter
0x01: Not
used (1)
0x07: 7
0x0F: 15
0x1F: 31
R/W
0x0214
0x0294
0x0314
0x0394
Moving
average
Range 2^n
(n =
0...16)
7
OPERATION