Connection
Dimension
Induction air ejector
DN 6/8
Span gas inlet
Clamping ring screw connection (hose fitting)
4 mm inner diameter
6 mm outer diameter
Gas outlet
DN 8/10
13.2.8
Sample gas conditions
Table 21: Sample gas characteristics
Sample gas at sampling point
Characteristic
Process temperature
10 ... 550 °C
Sample gas temperature assembly:
•
Sample gas probe
•
Sample gas line
•
Cell
Temperature:
•
Approx. 200 °C
•
Approx. 200 °C
•
Approx. 200 °C
Process pressure
–200 ... +200 hPa relative
Dust load
< 200 mg/m
3
13.2.9
Connections in analyzer
13.2.9.1
Power supply - connection / fuses
Overview
The power supply is located on the left on the analyzer.
UPS
Power supply - connections
Table 22: Power supply - connections
Name
Supply
MAINS UPS (3-pole)
Power supply for electronics unit (internal)
MAINS (5-pole)
External power supply
F1
Internal
F2
Internal
Table 23: Connection terminal - power voltage connection on the analyzer
Wire
Cross-section in
mm
2
Cross-section in
AWG
Tightening torque Nm
rigid
0.75 ... 10.0
18 ... 8
1.2 ... 1.5
flexible with ferrules
0.5 ... 6.0
18 ... 8
flexible with ferrules with insu‐
lating collar
0.5 ... 6.0
18 ... 8
13
TECHNICAL DATA
58
O P E R A T I N G I N S T R U C T I O N S | MCS200HW-MP
8026608/AE00/V1-0/2021-10 | SICK