Heating temperature of all parts with sample gas contact: 200 °C
•
Pump: Ejector pump in cell (operated with instrument air)
•
The analysis system uses status indicators to signal the current operating state:
•
The analysis system switches to operating state “System Stop” automatically when
a malfunction occurs
“System Stop” corresponds to classification “Failure”:
°
The sample gas line and the sample gas path in the analyzer are automati‐
cally purged with instrument air in this mode.
°
Measured values are updated further.
Check (validation) and adjustment
•
Zero point adjustment
•
Reference point adjustment
•
Adjustment with internal adjustment filter
Operating using the display
It is possible to operate the device using the display.
Operation via external PC (optional)
Operator menus and measured value displays are also available for easy use on an
external PC via the Ethernet connection (with Google Chrome browser and SOPASair).
3.3.2
Installation plate
Overview
On the installation plate:
•
Control unit
•
Measurement technology
•
Analog and digital interfaces
PRODUCT DESCRIPTION